MODEL 9DP-1 Operator’s Manual
Ludlum Measurements, Inc. 42 December 2015
Note: The error associated with displaying the integrated dose on the
measurement view is one numerical unit. As a result, an alarm value of 200
uR may appear to generate an alarm when the integrated dose is between
199.9 and 200.1 uR; however, the actual detected integrated dose was 200.0
uR.
Enable
“Enable” represents the ability to enable the integrated dose alarm for the
current view. “Enable” is a special field and valid values include “Off”
and “On.”
Value
“Value” represents the numerical value of the integrated dose alarm given
the supplied multiplier and units. “Value” is a numeric field with valid
values from 0 to 999.
Multiplier
“Multiplier” represents the multiplier associated with the integrated dose
alarm. “Multiplier” is a special field with valid values of micro, milli, unity,
kilo, and mega.
Units
“Units” represents the units associated with the radiation rate alarm.
“Units” is a non-editable field as it must be equal to the units associated
with the current view measurement rate.
6.5.3 View 2: Sv/h
“View 2: Sv/h” is a sub-menu allowing access to the “Appearance,” “Functions,”
“Alerts,” and “Alarms” associated with “View 2” Sv/h”. The menu features are
identical to the menu features for view 1 (see section 6.5.2).
6.5.4 View 3: Gy/h
“View 3: Gy/h” is a sub-menu allowing access to the “Appearance”,
“Functions,” “Alerts,” and “Alarms” associated with “View 3: Gy/h”. The menu
features are identical to the menu features for View 1 (see section 6.5.2).
6.5.5 View 4:
“View 4:” is a sub-menu allowing access to the “Appearance,” “Functions,”
“Alerts,” and “Alarms” associated with “View 4:”. The menu features are
identical to the menu features for View 1 (see section 6.5.2).