Masimo LiDCO Chapter 4: Operation
www.masimo.com 31 Masimo
Upper limit that triggers an
alarm
Off, or 42-259 in
steps of 1.
Lower limit that triggers an
alarm
Off, or 41 to 258 in
steps of 1.
The High Limit is upper
threshold that triggers an
alarm.
32-199 in steps of 1,
or Off
When set to Off,
alarm is disabled.
Low Limit is the lower
threshold that triggers an
alarm.
31-198 in steps of 1,
or Off
When set to Off,
alarm is disabled.
The High Limit is upper
threshold that triggers an
alarm.
37-219 in steps of 1,
or Off
When set to Off,
alarm is disabled.
Low Limit is the lower
threshold that triggers an
alarm.
36-218 in steps of 1,
or Off
When set to Off,
alarm is disabled.
Value below High Limit to
trigger caution light
Off, or 1 to 28 in
steps of 1.
Value above Low Limit to
trigger caution light
Off, or 1 to 28 in
steps of 1.
Length of time that the audible
alarm remains silenced
* Masimo LiDCO has the ability to control alarm priority based on the clinical use case. When Audible
Alarms are set to Off, the alarm priority is Low. When Audible Alarms are set to On, the alarm priority is
Medium.
BP Trends
Highest value that can be displayed
Lowest value that can be displayed