Masimo LiDCO Chapter 4: Operation
www.masimo.com 31 Masimo
Options
Description
Alarm
Priority
Factory
Default
Selectable Settings
Systolic High
Limit
Upper limit that triggers an
alarm
Low/Medium*
Off
Off, or 42-259 in
steps of 1.
Systolic Low
Limit
Lower limit that triggers an
alarm
Low/Medium*
Off
Off, or 41 to 258 in
steps of 1.
Diastolic High
Limit
The High Limit is upper
threshold that triggers an
alarm.
Medium
Off
32-199 in steps of 1,
or Off
When set to Off,
alarm is disabled.
Diastolic Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium
Off
31-198 in steps of 1,
or Off
When set to Off,
alarm is disabled.
MAP High Limit
The High Limit is upper
threshold that triggers an
alarm.
Medium
105
37-219 in steps of 1,
or Off
When set to Off,
alarm is disabled.
MAP Low Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium
70
36-218 in steps of 1,
or Off
When set to Off,
alarm is disabled.
MAP High
Caution Range
Value below High Limit to
trigger caution light
NA
5
Off, or 1 to 28 in
steps of 1.
MAP Low
Caution Range
Value above Low Limit to
trigger caution light
NA
5
Off, or 1 to 28 in
steps of 1.
Silence
Duration
Length of time that the audible
alarm remains silenced
NA
2 min
30 sec, 1, 2, or 5
min.
Audible
Alarms*
Disables audible alarm
NA
Off
On or Off
* Masimo LiDCO has the ability to control alarm priority based on the clinical use case. When Audible
Alarms are set to Off, the alarm priority is Low. When Audible Alarms are set to On, the alarm priority is
Medium.
BP Trends
Options
Description
Factory Default
Selectable Settings
Y-Axis Max
Highest value that can be displayed
180
25 to 300 in steps of 5
Y-Axis Min
Lowest value that can be displayed
30
20 to 295 in steps of 5
Heart Rate (HR)
Alarms