Masimo LiDCO Chapter 4: Operation
www.masimo.com 36 Masimo
Oxygen Consumption Index (VO2i)
Oxygen Consumption Index (VO2i) will display only when Index is selected for Oxygen Consumption
parameter display.
Alarms
Options
Description
Alarm
Priority
Factory
Default
Selectable
Settings
High Limit
Upper limit that triggers an alarm
Low/Medium*
Off
Off, or 180-450 in
steps of 10.
Low Limit
Lower limit that triggers an alarm
Low/Medium*
Off
Off, or 110-400 in
steps of 10.
High Caution
Range
Value below High Limit to trigger
caution light
NA
Off
Off, or 10-40 in
steps of 10.
Low Caution
Range
Value above Low Limit to trigger
caution light
NA
Off
Off, or 10-40 in
steps of 10.
Silence
Duration
Length of time that the audible
alarm remains silenced
NA
2 min
Off, or 30 sec, 1, 2,
and 5 min
Audible
Alarms*
Disables audible alarm
NA
Off
On or Off.
Trends
Options
Description
Factory Default
Selectable Settings
Y-Axis Max
Highest VO2i value that can be displayed
300
100-3000 in steps of 100
Y-Axis Min
Lowest VO2i value that can be displayed
0
0-2900 in steps of 100
Oxygen Delivery (DO
2
)
Alarms
Options
Description
Alarm
Priority
Factory
Default
Selectable Settings
High Limit
Upper limit that triggers an
alarm
Low/Medium*
1100
Off, or 350-1500 in
steps of 50.
Low Limit
Lower limit that triggers an
alarm
Low/Medium*
550
Off, or 300 to 1000 in
steps of 50.
High Caution
Range
Value below High Limit to
trigger caution light
NA
100
Off, or 50 to 300 in
steps of 50.