13 Malfunctions
73
Incident Possible causes What you can do
193 The left-hand chamber of the internal storage
tank (waste water) must be empty for the
pending rinsing process.
NOTICE!Danger of draining through the
emergency overflow. Check the water level
and empty the left-hand chamber of the
internal storage tank (waste water)
completely.
203 No log output options have been set. Check the configuration in the menu
"Settings" > "Logging".
204 The internal log memory is full. Output the log saved in the steam sterilizer
on any output medium or adapt the general
output options in the “Settings” > “Logging”
menu.
207 see event 203
208 see event 204
211 see event 204
214 The steam sterilizer has not recognized the
CF card; it cannot be read, or it is full.
1. Check whether the CF card has been
inserted correctly (do not insert under
voltage).
2. Make sure that the CF card is not larger
than 4 GB.
3. Check whether the write-protection has
been set on the CF card by mistake.
4. Test the CF card on a computer.
5. Check whether the memory on the CF is
full. If the memory is full, transfer the log files
from the CF card to another computer and
delete the files on the CF card.
6. Transfer the present log files from the CF
card to a computer and re-format the CF
card in the steam sterilizer.
7. The CF card is defective or incompatible It
is possible that a non-MELAG CF card has
been used.
Please note: We recommend using only
original MELAG CF cards.
218 The attempt was made to overwrite a write-
protected log with a log of the same name.
1. Transfer the log file the CF card to another
computer and delete the file from the CF
card.
2. Insert the empty CF card in the card slot
and enter the log again.
221 The CF card or a sub-directory of the CF
card is full.
1. Transfer the present log files from the CF
card to a computer.
2. Re-format the CF card in the steam
sterilizer.
223 The CF card has not been recognized. 1. Transfer the present log files from the CF
card to a computer and format the CF card in
the steam sterilizer.
2. Try again.
224 see event 223
228 see event 223
229 The CF card was removed from the slot
during a writing/reading action.
Never remove the CF card from the slot
whilst it is being written or read. Insert the CF
card in the card slot again and repeat the
procedure.