EasyManua.ls Logo

Metek 933 - Analog Output Calibration; Current Calibration

Default Icon
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-4 | Model 933 UV Analyzer
Analog Output Calibration
Current Calibration
The current outputs are calibrated at the factory. If a current output mod-
ule is replaced or added, calibrate that output (from the Output Setup
dialog box). Calibration of a current output is performed by entering the
measured low-scale (Zero) and full-scale (Span) signals for each output
(Figure 5-4). These values are used to offset the output to the correct val-
ues.
When performing current or voltage calibration, do not enable (check)
V/I Low Limit 4 mA until after calibration is complete.
To calibrate the current outputs:
1. Measure the Zero (low-scale) signal of each output:
a. Click Zero Cal to change its status to On. The status of Normal, Mid-
Scale, and Span Cal should be Off.
b. Connect a current meter to Pins 1 and 2 (at J109 on the Customer
I/O board) and measure the current of Output 1. Under Current
Outputs, enter the value next to Zero (mA), under the first column
(Output 1). Place a 250 Ohm resistor in series of the multi-meter for
accurate calibration.
Repeat this step for Output 2 (Pins 3 and 4), Output 3 (Pins 5 and 6),
and Output 4 (Pins 7 and 8), recording the value for each output in
the corresponding field.
NOTE
Figure 5-4.
Output Setup dialog
box.
Setup (tab)Output

Table of Contents

Related product manuals