EasyManua.ls Logo

Metek 933 - Customer Connections; Analog Outputs; Digital Communication; Digital Inputs

Default Icon
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-4 | Model 933 UV Analyzer
Customer Connections
Analog Outputs
4 isolated 4–20 mA
DC
, loop- or self-powered, 30 V
DC
maximum.
4 non-isolated 1–5 V
DC
.
5 independent sets of SPDT, Form C, potential free relay contacts,
2 A maximum at 240 V
AC
.
Digital Communication
Modbus
®
port: RS-485
Service port: RS-232/RS-485
Digital Inputs
For Auto-Cal and Alarm Enable/Clear:
0–24 V
DC
, 30 V
DC
Maximum.
Electrical Requirements
Maximum Start-Up Power
< 500 W maximum start-up with continuous average, depending on am-
bient temperature (excluding Sample and Vent Line).
Supply Voltage / Maximum Current
ATEX/Hazardous Locations:
120 V
AC
, 50/60 Hz, 3 A Maximum
240 V
AC
, 50/60 Hz, 1.5 A Maximum
CSA:
104~132 V
AC
, 50/60 Hz, 3 A Maximum
208~264 V
AC
, 50/60 Hz, 2 A Maximum
Sample Gas Flow Rate
Maximum 20 NLM (42.4 SCFH) for the High Pressure/Low Range Kit.

Table of Contents

Related product manuals