Calibration | 5-7
c. Click Apply to save the changes.
d. To test the accuracy of the output channels, click Mid-Scale to turn
On this function. Verify the voltage outputs are 3 V, ± 5 mV.
e. Click Span Cal to change its status to Off.
3. Click Normal to change its status to On, which activates the outputs.
The outputs will now track the concentration readings.
4. Click OK then Apply. Save the change to EEPROM.