EasyManua.ls Logo

Metek 933 - Page 207

Default Icon
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Troubleshooting | 6-63
Alarm Type / Alarm Condition Description and Suggested Corrective Action
Warning ALC One or both lamp pulse current-control signals has exceeded 9500 mV. The
Lamp Maximum range is typically 4500–7800 mV. When this alarm occurs,
the analyzer will initiate an Auto-Zero and then an Auto-Setup, to a maximum
of three times.
If the lamps are within range after the Auto-Setup, the counter resets to zero.
If not, the alarm will remain on to indicate there is a problem with the lamps,
which may need replacing.
This alarm may occur during a cold start, but should clear after the source
lamps have stabilized.
Corrective Action:
Take appropriate safety precautions, open the Upper Enclosure, and:
Check that the lamp sockets are attached securely to the lamps and that
the Lamp Retaining Bracket is in place.
Check that the lamps are inserted completely into the Lens Block in the
Detector Block Assembly.
Check the at cables between the Optical Bench board and Optical
Bench for proper connections and inspect them for damage (cuts, nicks,
burn marks, etc.).
The light levels may be too low as a result of natural aging of the lamps.
This is an indication that the lamps are nearing the end of their useful lives
and should be replaced.
Eventually, a “Fault Lamp Failure” alarms will be activated. Refer to the
description of this alarm for more details.
As a temporary solution, initiate an Auto-Setup to increase the PMT
gain to compensate for the reduced light levels (see “The Auto-Setup
Sequence” in this chapter).
Setup (tab)Optical
Bench

Table of Contents