6-68 | Model 933 UV Analyzer
Alarm Type / Alarm Condition Description and Suggested Corrective Action
Fault Comm The Microcontroller board cannot establish reliable communications with
the Host Controller board.
Corrective Action:
Take appropriate safety precautions and:
Cycle the analyzer AC power o󰀨 for a minimum of 15 seconds before
cycling it on again.
If this doesn’t correct the alarm, take appropriate safety precautions, click
Reset M933 on the Analyzer Information dialog box.
If this doesn’t correct the alarm, take appropriate safety precautions, open
the Electronics Enclosures, and:
- Reset the Host Controller board (SW100, Lower Enclosure – see
Figure 6-15).
- Reset the Microcontroller board (SW400, Upper Enclosure – see
Figure 6-16).
Check the connections of the cable between the Host Controller (J300)
and Micro-Interface (J104) boards. Inspect it for damage (cuts, nicks,
burn marks, etc.).
Replace the Microcontroller, Host Controller, and/or Micro-Interface
boards. Contact AMETEK to verify operation before removing any of
these boards.