5 Detailed description
774 Oven Sample Processor, Instructions for use
124
5.11.3.14 Info.ActualInfo.Inputs.Status read only
Status of the Input lines (Input0…7) of the Remote interface.
$Q sends the signal state as decimal number e.g.
10 00001010 binary 2
1
+ 2
3
Input1 and Input3 active
(active = low, inactive = high)
See also page 121ff.
5.11.3.15 Info.ActualInfo.Outputs.Status read only
Status of the output lines (Output0…13) of Remote interface. See 5.11.3.14.
5.11.3.16 Info.ActualInfo.Display.L1 read only
Info.ActualInfo.Display.L2 read only
Text of the first and second line of the LCD display.
5.11.3.17 Info.ActualInfo.Counter.Sample read only
Info.ActualInfo.Counter.Maximum read only
Currently processed sample number and total number of samples.