5.3 Menu organization
774 Oven Sample Processor, Instructions for use
73
Parameter
>dosing unit def.
Submenu for dosing unit settings
Open the submenu with <ENTER>
>dosing unit def.
dosing unit 1
Select dosing unit
1…12
After entering the address of the dosing unit and confirming with <ENTER> it
will be displayed in the first menu line.
>dosing unit 1
dos.rate: 160 ml/min
Set the dosing speed
0.01…160 ml/min,
max.
>dosing unit 1
fill.rate: 160 ml/min
Set the filling speed
0.01…160 ml/min,
max.
The following entries are only valid for 700 Dosinos. For details about Dosinos
and dosing units, see page 92ff.
>dosing unit 1
dosing port 1
Define dosing outlet
1…4
>dosing unit 1
filling port 2
Define filling inlet
1…2…4
>dosing unit 1
rinsing port 2
Define rinsing inlet (when changing
dosing units)
1…2…4
>dosing unit 1
preparation port 2
Define inlet for the preparation cycle
1…2…4
>dosing unit 1
drain port 4
Define air inlet for emptying
1…4