5 Detailed description
774 Oven Sample Processor, Instructions for use
74
parameters
>manual stop
Submenu for defining reaction to manual stop
Open the submenu with <ENTER>
The following entries define the commands to be executed when the
<STOP> key is pressed. This allows connected peripherals to be stopped
automatically or the functions of the 774 Oven Sample Processor to be reset.
>manual stop
Rmt CTL: ********
Signal output via remote interface
STOP device1,
14 bit (1,0 or *)
>manual stop
RS232 CTL:
Data to be transmitted via RS 232 in-
terface
14 ASCII characters Clear value '&M;$S'
>manual stop
FLOW: cont.
Leave gas flow unchanged or
deactivate it
cont., OFF
>manual stop
HEATER: INIT °C
Set oven heater to specific tempera-
ture or switch it off
INIT, OFF,
50…250 °C