5.6 ORP parameters
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
Setting Meaning Selection Default value
P2.6
CALIBRATION DATA
ORP calibration data VIEW | PRINT
P2.7
SET DUE CAL
ORP calibration time NO | HOURS | DAYS NO
P2.8
RESET SETTINGS
Resetting ORP settings YES | NO NO
P2.9
TEMPERATURE CAL
ORP temperature off-
set
YES | NO NO
P2.6 ORP calibration data
Information about the last calibration performed is displayed or printed
under P2.6.
VIEW
The following results are shown automatically on the display, one after the
other:
Date and time of calibration
Offset value of electrodes in mV
Calibration temperature
PRINT
The following results are printed out:
Model number and serial number
Date and time of calibration
Offset value of electrodes in mV
Calibration temperature
NOTE
Use only the original printer specified by the manufacturer.
Ensure that the printer is connected and switched on.
Ensure that the paper roll and the cassette are inserted correctly.
Additional information can be found in the manual for the printer.
P2.7 ORP calibration time
P2.7 is used to set the time until the next calibration. The P2.7 setting is
important for GLP protocols. No default value is set for the calibration
time.