5.3 Temperature measurement
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
Operation mode Meaning
The device is in set-up mode. The configuration
menus can relate to the properties of the
parameters, the data processing settings or the
device settings.
The device is in retrieve memory mode. The
saved values of the respective parameter are
displayed.
Changing the operation mode
After switching on, the device is automatically in measuring mode.
1
Press the button to switch between measuring mode and cali-
bration mode, respectively.
2
Press the button in measuring mode to switch to set-up mode.
3
In measuring mode, press the button to switch to retrieve mem-
ory mode for the required parameter.
The most recently saved values appear on the display.
5.3 Temperature measurement
Automatic temperature compensation (ATC)
The sample temperature is measured directly using a Pt1000 temperature
sensor. The temperature sensor can either be external or integrated in the
electrode.
Manual temperature compensation (MTC)
If no temperature sensor is connected, then the temperature value has to
be changed manually:
Setting the temperature manually
1
Keep the or buttons pressed for 3 seconds until the temp-
erature value begins to flash.