7.3 Messages
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
Problem Cause Remedy
A non-Metrohm power
supply unit is connected.
Use only the power supply unit provided
during measurement operation.
The measured value
is evidently wrong.
The pH calibration is faulty. Check and repeat the calibration.
Check and/or replace the buffers.
Check the buffer selection in the settings.
The temperature specifica-
tion is incorrect.
Enter the correct measuring temperature.
The glass membrane or the
diaphragm is contam-
inated.
Clean the membrane or the diaphragm fol-
lowing the instructions in the correspond-
ing leaflet.
The electrolyte is overaged. Replace the electrolyte.
The sensor is defective. Replace the sensor.
The slope is insuffi-
cient during calibra-
tion.
The glass membrane or the
diaphragm is contam-
inated.
Clean the electrode following the instruc-
tions in the leaflet.
The hydrated layer of the
glass membrane is reduced
by anhydrous solutions.
Immerse the glass membrane in deionized
water for 5 min to restore the hydrated
layer.
The buffer solutions are not
OK.
Replace the buffer solutions.
The sensor is "worn out". Replace the sensor.
7.3 Messages
The following error messages may appear during calibration:
Error message
Meaning
NOT STABLE
The button was pressed during an unsta-
ble signal. Wait for the stability signal to
confirm the calibration point.
WRONG BUFFER The buffer is contaminated or is not a member
of one of the detected buffer families.
SLOPE OUT OF RANGE The slope of the calibration curve is outside the
accepted range between 80% and 120%.