5.7 Data processing
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
3
Saving data manually
Press the button as soon as a value needs to be saved.
5.7.5 Printing data
Example: Printing out the pH value and calibration data manually.
1
Opening the settings
Use the button to change to the range.
Open the P8.0 memory settings.
Use the and buttons to select the P8.1 memory setting.
Confirm the selection with the
button.
2
Selecting Manual data collection
Use the and buttons to select MANUAL.
Confirm the selection with the button.
3
Selecting printing
Select the P8.3 memory setting.
Confirm the selection with the button.
Use the and buttons to select PRINTER.
Confirm the selection with the
button.
4
Selecting the print format
Select the P8.4 memory setting.
Confirm the selection with the button.
Use the and buttons to select COMPLETE.
Confirm the selection with the
button.
5
Opening the measuring mode
Press the button to return to measuring mode.
6
Printing data
Press the button to print the calibration data in GLP format.