5.7 Data processing
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
Default value: SIMPLE
NOTE
The data is printed out in the following cases:
■ As soon as a printout of certain data for a certain parameter is
started for the first time.
■ As soon as a new calibration is performed.
5.7.3 Automatic data collection
Example: Automatically record the pH value in the internal device memory
every 2 minutes.
1
Opening the settings
■ Use the button to change to the range.
■ Press the button to access the settings.
■ Open the P8.0 memory settings.
■ Confirm the selection with the
button.
■ Use the and buttons to select the P8.1 memory setting.
■ Confirm the selection with the
button.
2
Setting the frequency range
■ Use the and buttons to select MINUTES.
■ Confirm the selection with the button.
■ Use the
and buttons to change the flashing number on
the display to 2.
■ Confirm the setting with the button.
3
Opening the measuring mode
■ Press the button twice to return to measuring mode.
The symbol appears at the bottom of the display and shows that
an automatic data collection in a frequency range has been set.
4
Starting the data collection
Press the button to start the data collection.