Seite 101
Appendix | Optional Accessories
thermoMETER CTL
A 1.3 Water Cooled Housing
Fig. 35 Dimensions water cooled housing [TM-W-CTL], hose connection: 6x8 mm, thread (fitting): G 1/8 inch
Dimensions in mm, not to scale
i
To avoid condensation on the optics an air purge collar is recommended.
Water flow rate: approx. 2 l/ min
i
Cooling water temperature should not exceed 30 °C.
The sensor can be used at ambient temperatures up to 85 °C without cooling. For applications, where the
ambient temperature can reach higher values, the usage of the optional water cooled housing is recommend-
ed (ambient temperature up to 175 °C). The sensor should be equipped with the optional high temperature
cable (ambient temperature up to 180 °C).