EasyManua.ls Logo

MICRO-EPSILON thermoMETER CTL Series - Entering of the Calibration Code

MICRO-EPSILON thermoMETER CTL Series
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 49
Electrical Installation
thermoMETER CTL
7.5.1 Entering of the Calibration Code
Every sensor has a specific calibration code, that is clearly printed on a label on the sensor, see Fig. 14.
Fig. 14 Calibration code
i
Please do not remove this label or make sure the code is noted anywhere. The code is needed if the
sensor has to be exchanged.
Every sensor has a specific calibration code, which is printed on the sensor. For a correct temperature mea-
surement and functionality of the sensor this calibration code must be stored into the controller.
The calibration code consists of five blocks with 4 characters each.
Example: EKJ0 - 0OUD - 0A1B A17U 93OZ
block 1 block 2 block 3 block 4 block 5
i
After exchanging a sensor the calibration code of the new sensor must be entered into the controller.
For entering the code, please press the
and -Taste (keep pressed) and then the key, see Fig. 30.
The display shows HCODE and then the 4 signs of the first block. With
and each sign can be changed.
switches to the next sign or next block.

Table of Contents