7
Pre Installation
Installation Requirements
Feedwater
Requirements
Type of Feedwater Potable
Flowrate
≥ 40 LPH
Minimum Pressure
≥ 0.5 bar
Maximum Pressure
≤ 6 bar
Conductivity
< 2000 µS/cm
Temperature
5 to 35 °C
pH 4 - 10
Fouling Index < 10
Iron < 0.1 ppm as CaCO
3
Aluminum < 0.05 ppm as CaCO
3
Manganese < 0.05 ppm as CaCO
3
Free Chlorine < 1 ppm
Langelier Saturation Index < + 0.2
TOC < 2000 ppb
Feedwater
Connection Needed
Feedwater Piping Connection 1/2 inch Male GAZ, NPTM or BSPM
Reject Flow
Requirement
Drain Capacity 30 LPH
Environmental
Requirements
Indoor Use Only
Storage Temperature
5 °C < T < 40 °C
Operating Temperature
°
°
Relative Humidity
Should not exceed 80% for temperatures below 31 °C
Should stay within 50% to 80% between 31 °C and 40 °C.
Altitude < 3000 metres
Installation Category II
Pollution Degree 2