EasyManuals Logo

Millipore direct-q 3 R User Manual

Millipore direct-q 3 R
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
Installation
28
How to Show Resistivity or Conductivity Units (C01)
With Temperature Compensated or non Temperature Compensated values
Temperature compensation is a way of ‘standardizing’ Resistivity or Conductivity to measurements that would be
seen if the water temperature was 25 °C.
1
Press the Main and “+“ Buttons together
to enter the menu. The Display will show
C01 and the units chosen. The following
Display shows Temperature Compensated
Resistivity Units: M.cm @25 °C.
2
Press the “+” or “- Button to select
Non Temperature Compensated
Resistivity Units: M.cm.
3
Press the “+” or “- Button to select
Temperature Compensated Conductivity
Units: µS/ cm @25 °C.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Millipore direct-q 3 R and is the answer not in the manual?

Millipore direct-q 3 R Specifications

General IconGeneral
BrandMillipore
Modeldirect-q 3 R
CategoryWater Filtration Systems
LanguageEnglish

Related product manuals