69
Troubleshooting
Problem Possible causes What to do
The Display screen is blank.
The power cord is not plugged in.
No source of electrical power.
Main Power Fuse is blown.
Check that the power cord is
plugged in.
Check the source of electrical
power.
Contact your technical support.
In FILLING TANK mode, the
Tank Level display is not
rising.
(The Tank Level display
should be rising steadily.)
The Tank Outlet Valve is open. The
water in the tank is diverted into
the drain. No water is kept in the
tank.
The tank level is not calibrated
properly.
The RO membrane is dirty or
clogged.
Close the Tank Outlet Valve.
Check the Feedwater source.
See
Section How to Calibrate the
Tank Level (C04).
Replace the SmartPak cartridge.
Low water flow or no water
flow when the Dispensing
Button is pressed.
The water level is less than 10%.
No dispensing is allowed.
The Tank Outlet Valve is open. The
tank is not filling up with water.
Millipak Filter is air locked.
Final Filter is clogged.
Allow the system to go into
FILLING TANK mode until there is
enough water available in the tank,
viewed by the Tank Level display.
Close the Tank Outlet Valve.
Vent all air from the Millipak Filter.
Replace the Final Filter, see
Section
How to Replace the Final Filter.
DISPENSING mode
continuously runs even
when the Tank Level display
is below 10%.
The water level is not calibrated.
The system does not detect the
minimum level to disable
DISPENSING mode.
Calibrate the tank level. See
Section How to Calibrate the Tank
Level (C04).
AUTO-DISPENSING mode is
not accurate.
The Product Water flowrate is not
calibrated.
The Final Filter is clogged or
airlocked.
Calibrate the Product Water
flowrate. See
Section How to
Calibrate the Flowrate from the
Direct-Q R
(F02)
.
Replace the Final Filter.
The last RO Permeate
conductivity value is not
displayed when you press
the “+” button.
The measurement was not made
while dispensing water during
FILLING TANK mode.
The value is out of measurement
range.
Dispense water during FILLING
TANK mode to start a RO Permeate
conductivity reading again.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Millipore direct-q 3 R and is the answer not in the manual?

Millipore direct-q 3 R Specifications

General IconGeneral
BrandMillipore
Modeldirect-q 3 R
CategoryWater Filtration Systems
LanguageEnglish

Related product manuals