Direct-Q R Product Water Specifications
Water Flowrate
Specifications
Direct-Q R
System Type
RO water
flowrate
Dispensing
flowrate
Reject water
flowrate
(5°C < T < 35°C)
Direct-Q 3 R
3 LPH ± 15% @ 25°C
Up to 42 LPH 15 LPH to 25LPH
Direct-Q 5 R
5 LPH ± 15% @ 25°C
Up to 42 LPH 35 LPH
Direct-Q 8 R
8 LPH ± 15% @ 25°C
Up to 42 LPH 35 LPH
Product Water
Quality
Resistivity
18.2 MΩ.cm @ 25 °C
Conductivity
0.056 µS/cm @ 25 °C
Total Organic Carbon (TOC)
< 5 µg/L (ppb) (with UV)
< 10 µg/L (ppb) (without UV)
Micro-Organisms < 0.1 CFU/ml (with Millipak® Filter)