Installation
16
Installation of the SmartPak cartridge
ATTENTION
Open the Tank Outlet Valve before installing a new SmartPak cartridge. This
keeps the tank from filling until the SmartPak cartridge is rinsed out (A).
HAZARD Do not touch the UV Lamp when replacing the SmartPak cartridge.
Installation
A
Make sure the front cover is opened.
B
STANDBY should be viewed on the Display (B).
Remove the SmartPak cartridge from its shipping box.
Remove the protective caps on the ports of the SmartPak cartridge
and from the system.
C
Locate the O-rings on the ports (C). Wet them with water. It is
preferable to wet them with ultrapure water.