Maintenance
54
How to Empty the Tank (C03)
The tank can be fully emptied of water through the POU Dispenser. This option is used when performing a system
sanitization. Before emptying the tank, it is recommended to remove the Final Filter and to install the Barbed
Fitting and Tygon Tubing. Place the other end of the Tygon Tubing in a sink.
1
Press the Main and “+“ Buttons together to
enter the menu. The Display will show C01.
2
Press the Main Button 2 times. The Display
will show C03.
3
Press the “+” and “-“ Buttons together. The
system will dispense all the water in the tank
through the POU Dispenser. The Tank Level
display will be blinking.
NOTE:
To stop emptying the tank, press the
Dispensing Button.
4
When the system has finished emptying the
tank, the system will start FILLING TANK mode
automatically.