Maintenance
66
Install the Final Filter.
ATTENTION
If the SmartPak cartridge is not being replaced, then reinstall the old SmartPak
cartridge BEFORE powering ON the system. Otherwise, the system will go into
FLUSH mode for 15 minutes during which no Product Water will be available.
Close the front cover.
Plug the electrical cord to power ON the system.
Reset the UV Timer.
See
Section How to View or Reset the UV Lamp Timer (C05).