_____________________________________ TROUBLESHOOTING
STATUS / PROBLEM CAUSE REMEDY
EXCHANGE PACK. The operational lifetime of the Change the purification pack.
purification Pack has expired.
PACK OUT The Pak has not been installed Put the Pack back in place.
properly or has been removed.
EXCHANGE UV LAMP The operational lifetime of the Replace the UV lamp
UV lamp has expired.
CHECK UV LAMP Defective UV lamp Occurs when the UV lamp
doesnt light anymore.
EEPROM ERROR The unit needs to be initialised Press the OPERATE button twice
_____________________________________ TROUBLESHOOTING
STATUS / PROBLEM CAUSE REMEDY
EXCHANGE PACK. The operational lifetime of the Change the purification pack.
purification Pack has expired.
PACK OUT The Pak has not been installed Put the Pack back in place.
properly or has been removed.
EXCHANGE UV LAMP The operational lifetime of the Replace the UV lamp
UV lamp has expired.
CHECK UV LAMP Defective UV lamp Occurs when the UV lamp
doesnt light anymore.
EEPROM ERROR The unit needs to be initialised Press the OPERATE button twice
_____________________________________ TROUBLESHOOTING
STATUS / PROBLEM CAUSE REMEDY
EXCHANGE PACK. The operational lifetime of the Change the purification pack.
purification Pack has expired.
PACK OUT The Pak has not been installed Put the Pack back in place.
properly or has been removed.
EXCHANGE UV LAMP The operational lifetime of the Replace the UV lamp
UV lamp has expired.
CHECK UV LAMP Defective UV lamp Occurs when the UV lamp
doesnt light anymore.
EEPROM ERROR The unit needs to be initialised Press the OPERATE button twice