Environmental Characteristics Appendix
6 - 12 0070-10-0699-02 Accutorr V Operating Instructions
6.4 Environmental Characteristics
• Operating:
Temperature: 0°C to 40°C, 32°F to 104°F (Accutorr V &
Recorder)
10°C to 40°C, 50°F to 104°F (Predictive
Thermometer)
Humidity: 15 to 95% max, non-condensing.
Altitude: -1,250 to 10,000 feet (-381 to 3048 meters,
70.0 kPa to 106 kPa)
Shock and Vibration: Meets IEC 60068-2-27 for shock with peak
acceleration of 15.3g, 11 mSec duration, half
sine.
Meets EN 60068-2-64 for random vibration
with frequencies of 10 to 2000 Hz, resolution
of 10Hz, 10 minutes per axis, acceleration
spectral density of:
10 Hz to 100 Hz: 1.0 (m/s
2
)
2
/Hz
100 Hz to 200 Hz: -3 dB per octave
200 Hz to 2000 Hz: 0.5 (m/s
2
)
2
/Hz
Meets IEC 60068-2-6 for sinusoidal vibration
with Displacement / acceleration of 0.07 mm
or 1g, 58 – 62 Hz crossover frequency 10 Hz
to 500 Hz with 5 frequency sweep cycles at
each axial direction.
• Storage and Transportation:
Shipping: Meets ISTA Test Procedure 2A
(less than 100 lbs)
Temperature: -20°C to +60°C, -4°F to 140°F
Humidity: 10 to 95%, non-condensing
NOTE: The Accutorr V may not meet its performance
specifications if stored outside these temperature and
humidity specifications.
Altitude: -1,250 to 20,000 feet (-381 to 6096 meters,
46.6 kPa to 106 kPa)