Accutorr V Operating Instructions 0070-10-0699-02 1 - 5
General Description Recommended Test and Calibration Frequency
1.3 Recommended Test and Calibration Frequency
CHECK/MAINTENANCE ITEM FREQUENCY
Visual test When first installing or after reinstalling.
Power on test 1. When first installing or after reinstalling.
2. Following any maintenance or replacement of any main unit
part.
NIBP tests Accuracy test 1. If the user suspects that the measurement is incorrect.
2. Following any repairs or replacement of the NIBP module.
3. At least once per year.
Leakage test
Calibration
SpO
2
test
Temperature test
Analog output test If the user suspects that analog output is abnormal.
Bar code scanner test If the user suspects that bar code scan is incorrect.
Electrical safety tests Enclosure leakage
current test
1. Following any repair or replacement of the power module.
2. At least once every two years.
Earth leakage current
test
Patient leakage
current test
Patient auxiliary
current test
Recorder check Following any repair or replacement of the recorder.