Accutorr V Operating Instructions 0070-10-0699-02 ix
Notes Introduction
Notes
NOTE: The Accutorr V should be operated only by trained and 
qualified personnel.
NOTE: Use disposable and single use accessories only once.
NOTE: Place the equipment in a location where the screen can 
easily be seen and the operating controls can easily be 
accessed.
NOTE: In certain situations in which perfusion and signal strength 
are low, such as in patients with thick or pigmented skin, 
inaccurately low SpO
2
 readings will result. Verification of 
oxygenation should be made, especially in preterm infants 
and patients with chronic lung disease, before instituting 
any therapy or intervention.
NOTE: The instructions in this manual are based on the maximum 
configuration.
NOTE: The optional Temperature module kit must be installed only 
by trained personnel, and proper ESD prevention methods 
must be followed.
NOTE: Only devices specified by Mindray DS USA, Inc./Shenzhen Mindray 
Bio-Medical Electronics Co., Ltd shall be connected the RS-232 port.
NOTE: When the RS-232 connector is used for DIAP, barcode power 
must be set to OFF.
NOTE: Disconnect the Accutorr V from the mains to isolate it from 
the mains power during an emergency.
Safety Designations
Safety designations per IEC 60601-1 Standard:
Type of protection against electric shock Class 1 with internal electric power source. 
Where the integrity of the external protective 
earth (ground) in the installation or its 
conductors is in doubt, the equipment shall be 
operated from its internal electric power 
source.
Degree of protection against electric shock Monitor - Type B applied part.
NIBP - Type BF defibrillation protected 
applied part.
SpO
2
 - Type BF protected applied part.
Temp - Type BF protected applied part.
Supply Connection 100 – 240 VAC
50/60 Hz
0.85 – 0.5 A