Battery Maintenance and Replacement User Maintenance
4 - 6 0070-10-0699-02 Accutorr V Operating Instructions
4.5 Battery Maintenance and Replacement
4.5.1 Battery Maintenance
The Accutorr V uses a lithium ion battery. This battery type may be subject to local 
regulations regarding disposal. At the end of the battery life, dispose of the battery in 
accordance with any local regulations.
CAUTION: Recharge the Lithium ion battery while in the unit at room 
temperature. If using the Accutorr V in a hot environment, 
the Lithium ion battery may not charge when the unit is 
connected to AC.
CAUTION: Remove the battery if the Accutorr V is not likely to be used 
for an extended period of time.
CAUTION: Replace the Lithium ion battery with new battery if its run 
time is seriously diminished.
4.5.2 Battery Replacement
Use only the battery specified in Section 5.1.8.
1. Open battery compartment door, located on the bottom of the unit, by pressing the 
release tab.
2. Rotate the battery retaining clip away from the battery.
3. Slide out the old battery by pulling on the fabric tab attached to the battery.
4. With the main label facing the back of the Accutorr V, slide the replacement battery in 
until it clicks into place.
5. Rotate the battery retaining clip back over the battery.
6. Close the battery compartment door.
7. Batteries are shipped partially charged and require fully charging prior to use. Charge 
the Lithium ion battery the Accutorr V for 4.5 hours minimum prior to use.
NOTE: If the Accutorr V is plugged into the AC mains, the Battery 
Status Indicator (15) illuminates if a battery is inserted. The 
Battery Status Indicator also (15) illuminates if the Accutorr 
V is turned on with a battery inserted and the AC mains 
unplugged.