Automatic NIBP Measurements (Interval Mode) Operation
3 - 16 0070-10-0699-02 Accutorr V Operating Instructions
3.7.6 Automatic Retry
If an NIBP measurement fails, the Accutorr V retries measuring up to three times. The
message “NIBP RETRY” is displayed after the cuff deflates. If the measurement fails all three
retries, the LCD Display (17) shows the Sys, Dia, and Map values as “XXX” and PR as “–” if
SpO
2
not providing the pulse rate.