Accutorr V Operating Instructions 0070-10-0699-02 3 - 37
Operation Temperature Measurement
FIGURE 3-21 Probe Placement for Axillary Temperatures
FIGURE 3-22 Probe Placement for Axillary Temperatures
6. The monitor beeps when the temperature measurement is complete.
7. Withdraw the probe from the patient’s armpit.
8. Press the ejection button on the top of the probe to eject the probe cover.
9. Return the probe to the sheath.
3.12.5 Measuring Rectal Temperature
1. Verify that the rectal probe is connected to the probe connector, and the indicator lamp
illuminates to indicate that TEMP module is operating.
2. Select the temperature type.
3. Set TEMP POSITION to RECTAL (see Section 3.12.1).
4. Remove the probe from the probe sheath and insert it into a cover in the probe cover
box. Press the probe handle down firmly until the probe engages the cover.
5. After probe warms, insert the probe into the patient’s rectum. To insure proper tissue
contact, angle the probes lightly after insertion. Insertion depth is recommended at 1/2"
to 3/4" (1.25 to 2 cm) for adults and 1/4" to 1/2" (0.5 to 1.25 cm) for children. A
lubricant may be used if desired. The measurement will proceed similarly to the oral
measurement, and the final reading will be displayed when the display stops flashing.