Status and Error Codes Operation
3 - 56 0070-10-0699-02 Accutorr V Operating Instructions
NOTE: The prompt message PULSE SEARCH is for the DPM and
Nellcor SpO
2
modules only.
NOTE: The prompt message SPO2 MOTION is for the Nellcor SpO
2
module only.
STANDBY FAILED The unit failed to enter the standby mode. Following instructions in Section 3.4.1
again. If standby fails again, restart unit.
If failure continues, contact service
personnel.
SPO2 WEAK PULSE The SpO
2
signal is weak. Change the sensor site for better signals.
SPO2 MOTION SpO
2
sensor is moving. Eliminate sensor movement.
ALARM DISABLED! One or more alarms are switched off. Switch on alarm(s).
RECORDER BUSY The recorder is recording data. Wait till the end of the recording.
RESETTING… Module is resetting. Wait till resetting is finished.
CALIBRATING... The NIBP module is being calibrated. Wait till the end of the calibration.
CALIBRATION COMPLETE The calibration is finished. None
PNEUM TEST COMPLETE The test for air leakage is finished. None
RESET FAILED The NIBP module fails to be reset. None
TEMP WARMING UP TEMP module is warming up. Wait till TEMP module completes warm-
up.
PREDICTIVE TEMP READY TEMP module completes warming up
and predictive measurement can be
performed now.
Perform predictive TEMP measurement.
PREDICTIVE TEMP
COMPLETE
TEMP predictive measurement is finished. Check TEMP reading.
TEMP MEASURE
COMPLETE
TEMP monitoring is over Remove the TEMP probe from patient
and insert it in the probe sheath.
SERVER NOT EXIST Sever is not installed. Install the server.
CONFIGURATION
RESTORED
Last configuration is loaded successfully. None
FAC ADULT CONFIG
LOADED
ADULT factory configuration is loaded
successfully.
None
FAC PEDIATRIC CONFIG
LOADED
PEDIATRIC factory configuration is
loaded successfully.
None
FAC NEONATAL CONFIG
LOADED
NEONATAL factory configuration is
loaded successfully.
None
USER ADULT CONFIG
LOADED
ADULT user configuration is loaded
successfully.
None
USER PEDIATRIC CONFIG
LOADED
PEDIATRIC user configuration is loaded
successfully.
None
USER NEONATAL CONFIG
LOADED
NEONATAL user configuration is loaded
successfully.
None
PROMPT MESSAGE CAUSE ACTION