BeneLink Module Operator’s Manual  4-83 
Monitor output: alarms from Philips Respironics V60 (VRPT Protocol) Ventilator 
Priority  Label 
Technical alarms 
High 
O2 Supply Pressure Low 
Patient Disconnect 
High 
Technical 
alarms 
If the monitor shows the alarm [High Technical 
alarms], the external device has one or more of the 
following alarms: 
Low internal battery alarm status 
Primary alarm failure 
Air source fault alarm status 
O2 valve stuck closed alarm status 
Airway Obstructed? 
Check Expiration-Valve 
Medium 
Medium 
Technical 
alarms 
If the monitor shows the alarm [Medium Technical 
alarms], the external device has one or more of the 
alarm: High enclosure temperature. 
Low 
Airway Leak? 
Low 
Technical 
alarms 
If the monitor shows the alarm [Low Technical 
alarms], the external device has one or more of the 
alarm: Nonvolatile memory failure.