9-4 BeneLink Module Operator’s Manual
9.3.2 Setting the Serial Ports of the Ventilator
Brand Model Protocol Baud
Rate
Data
Bits
Stop
Bits
Parity Remark
Carefusion Vela / 115200 8 1 None /
Draeger Babylog
8000/8000plus
/ 9600 / 1 None
Cal. Config → config → com → Start bit:
1
Evita 2 Medibus 19200 / 1 Even Channel A: No need to set
Channel B: Menu Mode → Other
Evita 2 Dura Medibus 19200 / 1 Even
Cal. Conf. → Device → Interface
Evita 4 Medibus 19200 / 1 Even
Configuration → Ventilation → System
Defaults → Interface
Evita XL Medibus 19200 / 1 Even /
Infinity
V500/Babylog
VN500
Medibus,
Medibus X
19200 / 1 Even
System setup → system → interface →
COM → Start bit: 1
Savina 300 Medibus, 19200 / 1 Even
System Setup → Interface