EasyManua.ls Logo

Mindray BeneLink - Setting the Serial Ports of the Ventilator

Mindray BeneLink
227 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9-4 BeneLink Module Operators Manual
9.3.2 Setting the Serial Ports of the Ventilator
Brand Model Protocol Baud
Rate
Data
Bits
Stop
Bits
Parity Remark
Carefusion Vela / 115200 8 1 None /
Draeger Babylog
8000/8000plus
/ 9600 / 1 None
Cal. Config config com Start bit:
1
Evita 2 Medibus 19200 / 1 Even Channel A: No need to set
Channel B: Menu Mode Other
Evita 2 Dura Medibus 19200 / 1 Even
Cal. Conf. Device Interface
Evita 4 Medibus 19200 / 1 Even
Configuration Ventilation System
Defaults Interface
Evita XL Medibus 19200 / 1 Even /
Infinity
V500/Babylog
VN500
Medibus,
Medibus X
19200 / 1 Even
System setup system interface
COM Start bit: 1
Savina 300 Medibus, 19200 / 1 Even
System Setup Interface

Table of Contents