1-2  BeneLink Module Operator’s Manual 
1.2 Warnings 
WARNING 
 The BeneLink module receives parameter, alarm, and waveform data 
from external devices but will not modify this data. For example, the 
external device may display "---",  b u t  send "0" to the patient monitor. 
 There can be differences between the alarm priorities displayed on 
your patient monitors and the priorities displayed on external devices 
interfaced through BeneLink. For details please 
see "Output signals –
Alarms" section corresponding with each external device. 
1.3 Cautions 
CAUTION 
 First installation and debugging should be executed by Mindray
service personnel or authorized technician. 
 Please check the compatibility of the external device and the ID 
adapter before connection. Not doing so may may result in failure of
the two devices to work together as desired. 
 Ports on the BeneLink module are not conventional network
connectors. They are intended for connecting with the serial port of 
designated devices only. Do not connect them to public network 
interfaces. 
 The signal labels used on the patient monitor may be different from those 
given on the external device. For details please see "Output signals
–
Parameters" section corresponding with each external device. The alarms 
from external devices may be delayed before transmission to the patient 
monitor
.