BeneLink Module Operator’s Manual  9-11 
9.5 BeneLink Installation and Test Report 
Basic Information 
Hospital 
Department 
Serial number of ID adapter 
Nam
e of 
external device 
ID of the external device 
Type
 of serial 
port adapting 
cable 
Software version and other information of the 
external device 
Checking the connection 
Test Result 
(Yes/No) 
Can the patient monitor and the external device be assembled 
together using designated accessories? 
Does the green indicator of corresponding port on the BeneLink 
module illuminate while the other indicators are not?   
Are there numerics or characters displayed on the [Integrated 
Devices] screen of the patient monitor? 
Are the device type and parameter values displayed correctly on the 
[Integrated Devices] screen of the patient monitor when the 
external device just enters working mode? 
Is the ventilation mode correctly displayed on the patient monitor? 
Does it change correctly when the ventilation mode on the external 
device is changed?