9-2  BeneLink Module Operator’s Manual 
 Set the serial port of the external device by referring to 9.3 Setting the 
Serial Ports. 
 Set the pressure control mode and check if the ventilator or anesthesia 
machine works normally. 
3. Make sure the ID adapter is correctly configured, and the green indicator of 
corresponding port on the BeneLink module illuminates constantly. 
4. Access the [Integrated Devices] screen on the patient monitor. Check that the 
device type (ventilator or anesthesia machine) and ventilation mode are correctly 
displayed. 
5. Select parameters PEEP, Pmean, VTe, MV, I:E, and f(RR) respectively on the 
patient monitor and check if the parameter values displayed on the patient monitor 
are consistent with those displayed on the ventilator or anesthesia machine. 
6. Re-configure the above parameters on the ventilator or the anesthesia 
machine and check if the parameter values displayed on the patient monitor change 
accordingly. 
7. Trigger alarms [MV Low], [Airway Pressure Too High], [PAW High], [Peak High]
and [No Gas Supply] (no Air or O
2
) on the ventilator or the anesthesia machine. 
Check whether these alarm messages are correctly recorded in the alarm list of the 
patient monitor.   
8. Switch the ventilator or anesthesia machine to volume control ventilation 
mode. Check if the ventilation mode displayed on the patient monitor changes 
accordingly, and if the parameter values of PEEP, Pmean, VTe, MV, I:E, and f(RR) are 
correctly displayed.