EasyManuals Logo

Mindray BeneLink User Manual

Mindray BeneLink
227 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #184 background imageLoading...
Page #184 background image
9-2 BeneLink Module Operators Manual
Set the serial port of the external device by referring to 9.3 Setting the
Serial Ports.
Set the pressure control mode and check if the ventilator or anesthesia
machine works normally.
3. Make sure the ID adapter is correctly configured, and the green indicator of
corresponding port on the BeneLink module illuminates constantly.
4. Access the [Integrated Devices] screen on the patient monitor. Check that the
device type (ventilator or anesthesia machine) and ventilation mode are correctly
displayed.
5. Select parameters PEEP, Pmean, VTe, MV, I:E, and f(RR) respectively on the
patient monitor and check if the parameter values displayed on the patient monitor
are consistent with those displayed on the ventilator or anesthesia machine.
6. Re-configure the above parameters on the ventilator or the anesthesia
machine and check if the parameter values displayed on the patient monitor change
accordingly.
7. Trigger alarms [MV Low], [Airway Pressure Too High], [PAW High], [Peak High]
and [No Gas Supply] (no Air or O
2
) on the ventilator or the anesthesia machine.
Check whether these alarm messages are correctly recorded in the alarm list of the
patient monitor.
8. Switch the ventilator or anesthesia machine to volume control ventilation
mode. Check if the ventilation mode displayed on the patient monitor changes
accordingly, and if the parameter values of PEEP, Pmean, VTe, MV, I:E, and f(RR) are
correctly displayed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneLink and is the answer not in the manual?

Mindray BeneLink Specifications

General IconGeneral
BrandMindray
ModelBeneLink
CategoryControl Unit
LanguageEnglish