BeneLink Module Operators Manual 2-9
2. Insert the BeneLink module into the SMR or module rack.
3. Connect the ID adapter that matches the external device to the BeneLink
module with an RJ45 connecting cable.
4. Plug the ID adapter into the RS232 port on the external device. Some external
devices may have ports incompatible with the ID adapter. In this case, a serial
port adapting cable is required.
5. Adhere a device name label to the RJ45 connecting cable at the end close to the
BeneLink module. When the BeneLink module is connected to several external
devices, you can identify devices easily with these labels.
6. Switch on the external device.
After the external device is connected to the monitor, the indicators on both the ID
adapter and the BeneLink module illuminate to show that the monitor is successfully
communicating with the external device.