Mindray Consona N9 - Screen Display of Tissue Tracking QA

Mindray Consona N9
416 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Image Acquisition
Operator’s Manual 6 - 33
6.14.2 Screen Display of Tissue Tracking QA
1 Displays image used to generate
trace curve
/
2 Displays corresponding time of
AVO (aortic valve open)/AVC
(aortic valve close)/MVO
(mitral valve open)/MVC (mitral
valve close).
/
3 Display curve: Velocity/
Displacement/Strain/Strain Rate.
Each curve on the image is matched with a certain segment
in the cardiac segmentation model (6), identified by
different colors.
Velocity curve: X-axis represents time (s); Y-axis
represents velocity (cm/s).
Displacement curve: X-axis represents time (s); Y-axis
represents displacement (mm).
Strain curve: X-axis represents time (s); Y-axis
represents deformation of the tissue (%).
Strain-rate curve: X-axis represents time (s); Y-axis
represents strain by time (s
-1
).
4 Displays ECG trace /
5 Displays cardiac segmentation
model, and each segment name
is illustrated beneath the model.
In the figure, marks the peak position of the curve.
Under tracking status, click on a segment in the cardiac
segmentation model. The segment has “X” mark and its
corresponding calculating is eliminated.
Tap certain segment in the cardiac segmentation model,
the segment will turns grey and its corresponding curve
no longer displays.
You can get the current X/Y axis value by moving the
cursor onto one point on the curve; and if you press
<Set> at this time, the frame marker will move to the
spot.
The segment boundary color indicates the tracking quality.