7 - 38 Operator’s Manual
7 3D/4D
Tap [Current Plane] to reset the plane. The position of the current plane returns to the
initial condition. The angle of the current plane returns to 0°.
Tap [Current Plane] to reset the current plane. The position and the angle to the current
plane returns to the last MSP condition.
Other related Operations
Hiding the reference
Tap to hide or show the reference line.
Hiding the measurement
Tap to hide or show the measurement.
Switching between the single and quad-split display
Tap [Single]or [Quad] between the single and quad-split display.
7.16 Smart FLC (Smart Follicles Calculation)
CAUTION
Smart FLC result is provided for reference only, not for confirming a
diagnosis.
Smart FLC is not available in Smart 3D mode.
7.16.1 Basic Procedures for Smart FLC
Perform the following procedure:
1. Acquire necessary 3D/4D data and freeze.
2. Tap [Smart FLC] tab to enter Smart FLC, and the system is in the “Edit ROI” status ([Edit
ROI] button is highlighted).
3. Set the position and size of the ROI box.
4. Tap [Calc] on the touch screen.
5. Auto calculation is finished:
Section image (A, B, C) is enclosed with colored lines and the color corresponds to the
follicle in VR window.
3D image (VR);
Results on the right of the screen are displayed in a table.
V: Volume
d1, d2, d3: diameters of the follicle
MD: Follicle Diam
NOTE:
To ensure the correctness of the results, please select an image with clear follicle boundary when
entering smart FLC.