4  Setup
Operator’s Manual 4 - 11
2. Select or deselect the check box from the privilege list. 
3. Click [OK] to confirm the editing and exit the dialog box. 
The edited privileges will appear in the User List. 
Modify Passwords
The account password needs to be changed every 3 months.
The system administrator can modify all user passwords. The administrator password is empty by 
factory default. You can set this password.
An operator can only modify his/her own password. 
1. Select the user name to be modified in User List. 
2. Click [Change Password] to open the dialog box.
3. Enter current password, new password and confirm new password, then click [OK].
Configure Password Policy
Turn on the access control function and log in to the system as Administrator before you configure 
the password policy.
Click [Password Policy Config]:
Item Description Remark
Lockout 
Threshold
Set the maximum time that a user can input 
the wrong password. If you exceed the 
maximum times, your account will be locked. 
For example, assume that the 
“Lockout Threshold” is set to 
5, the “Reset Account 
Lockout Threshold after” is 
set to 60, and the “Lockout 
Duration” is set to 60. That is, 
a user inputs the wrong 
password for 5 times within 
60 minutes, the account is 
locked, and the user can log in 
to the system only after 60 
minutes. Other users with 
unlocked accounts can still 
log in to the system normally. 
Reset Account 
Lockout 
Threshold after
Set the duration allowed for a user to 
continuously input the wrong password. 
Lockout Duration Set the duration after an account is locked. 
Reset all lockout Reset all locked accounts.  /