Phantom Usage Illustration B-3
2. B1-B4: Lateral Resolution Target String Group,
The depth of this target string are at 10mm, 30mm, 50mm, 70mm for B1-B4 group respectively. The
horizontal distance of adjacent strings in each group are 4mm, 3mm, 2mm, 1mm respectively.
3. C: Dead Zone Target Group,
The adjacent target string lateral intervals are 10mm. The distance of the target strings to the
acoustic window are 4mm, 3mm, 2mm respectively.
4. D: Vertical Distance Target Group,
This target group includes 12 strings, with the adjacent string interval is 10mm.
5. E: Horizontal Distance Target Group,
This target group includes 6 strings, with adjacent string interval are 20mm.
6. F1-F3: Anechoic (Cyst-like target, diameter: 2 mm, 4mm, 6mm) Target Group,
Cyst-like cylindrical shape target, is imbedded in the TM material with diameter of 2mm, 4mm, 6mm.
Their axials are parallel to the target string, with location at the depth of 15mm, 30mm, 45mm.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-70 and is the answer not in the manual?

Mindray DC-70 Specifications

General IconGeneral
Power Supply100-240V AC, 50/60Hz
Imaging ModesB-mode, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ProbesConvex, Linear, Phased Array, Volume
ConnectivityUSB, Ethernet, VGA
Advanced Features3D/4D Imaging
ApplicationsAbdominal, Obstetrics, Gynecology, Cardiac, Small Parts, Vascular, Musculoskeletal, Urology
WeightApprox. 85 kg
Operating SystemWindows-based

Related product manuals