Passport Vâ„¢ Service Manual 0070-10-0705 3 - 47
Repair Information Troubleshooting Menus
CO
2
: Warming up
(DPM/Microstream only)
1The CO
2
sensor has not
reached its operating
temperature.
1It takes typically 30
seconds for the sensor to
warm up.
2Wait for the message to
go away.
CO
2
: Startup
(DPM only)
1The CO
2
is starting up. 1 Wait for the message to
go away.
Calibration Completed
Successfully
(DPM/Microstream only)
1Calibration finished and
successful.
/
Not Ready For Calibration
(DPM/Microstream only)
1Module in unable to
initialize calibration.
1Repeat calibration
procedure. If problem
persists, contact Technical
Support.
CO
2
: Zeroing...
(DPM/Microstream only)
1The device is adjusting
the sensor signal to
predefined range for
better measurement.
1This is normal operation.
Wait for message to
clear.
CO
2
: Zero Failed
(DPM/Microstream only)
1Signal cannot be adjusted
to predefined range
within zeroing duration.
1Check the CO
2
connections. After the
sensor's temperature
becomes stabilized,
perform a zero
calibration again. If
problem persists, contact
Technical Support.
Calibration Error
(DPM/Microstream only)
1The actual concentration
of calibration gas applied
is in wide discrepancy
with the concentration
value entered by the user
in the Calibration dialog.
1Make sure the
concentration value
entered is the same as the
calibration gas applied.
2Repeat calibration
procedure. If problem
persists, contact Technical
Support.
Caused by no gas or wrong
gas concentration
(Microstream only)
1No gas or wrong gas
concentration.
1Check the gas
connections or gas
concentration. Repeat
calibration procedure. If
problem persists, contact
Technical Support.
Caused by Measurement Error
(Microstream only)
1Measurement Error 1Repeat calibration
procedure. If problem
persists, contact Technical
Support.
Caused by no stable gas flow
(Microstream only)
1No stable gas flow 1Check the gas
connections or gas
concentration. Repeat
calibration procedure. If
problem persists, contact
Technical Support.
MESSAGE/ PROBLEM REASON SOLUTION

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish