V Series Operating Instructions 8 - 7
Alarms Alarm Limits
8.2.2.3 Temp Tab
TABLE 8-4 lists the alarm limit ranges for the parameters contained in the Temp tab.
8.2.2.4 Gases Tab
TABLE 8-5 lists the alarm limit ranges for the parameters contained in the Gases tab.
TABLE 8-4
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
Temperature ( T1-T3)
Unit of Measure: °F
Resolution: 0.1 °F
Invalid data: “- -
High Limit Off, 77.0 to 113.0
Default: Off
Off, 77.0 to 113.0
Default: Off
Off, 77.0 to 113.0
Default: Off
Low Limit Off, 59.0 to 104.0
Default: Off
Off, 59.0 to 104.0
Default: Off
Off, 59.0 to 104.0
Default: Off
Temperature ( T1-T3)
Unit of Measure: °C
Resolution: 0.1 °C
Invalid data: “- -
High Limit Off, 25.0 to 45.0
Default: Off
Off, 25.0 to 45.0
Default: Off
Off, 25.0 to 45.0
Default: Off
Low Limit Off, 15.0 to 40.0
Default: Off
Off, 15.0 to 40.0
Default: Off
Off, 15.0 to 40.0
Default: Off
Delta Temperature
Unit of Measure: °F
Resolution: 0.1 °F
Invalid data: “- -
High Limit Off, 1.0 to 10.0
Default: Off
Off, 1.0 to 10.0
Default: Off
Off, 1.0 to 10.0
Default: Off
Low Limit N/A N/A N/A
Delta Temperature
Unit of Measure: °C
Resolution: 0.1 °C
Invalid data: “- -
High Limit Off, 1.0 to 10.0
Default: Off
Off, 1.0 to 10.0
Default: Off
Off, 1.0 to 10.0
Default: Off
Low Limit N/A N/A N/A
TBlood
Unit of Measure: °F
Resolution: 0.1 °F
Invalid data: “- -
High Limit Off, 95.0 to 110.0
Default: Off
Off, 95.0 to 110.0
Default: Off
N/A
Low Limit Off, 80.0 to 100.0
Default: Off
Off, 80.0 to 100.0
Default: Off
N/A
TBlood
Unit of Measure: °C
Resolution: 0.1 °C
Invalid data: “- -
High Limit Off, 35.0 to 43.0
Default: Off
Off, 35.0 to 43.0
Default: Off
N/A
Low Limit Off, 26.0 to 38.0
Default: Off
Off, 26.0 to 38.0
Default: Off
N/A
TABLE 8-5
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
SpO2
Unit of Measure: %
Resolution: 1%
Invalid data: “- -
High Limit Off, 75 to 101
Default: 101
Off, 75 to 101
Default: 101
Off, 75 to 101
Default: 101
Low Limit 70 to 99
Default: 88
70 to 99
Default: 90
70 to 99
Default: 92

Table of Contents

Other manuals for Mindray V12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V12 and is the answer not in the manual?

Mindray V12 Specifications

General IconGeneral
BrandMindray
ModelV12
CategoryMedical Equipment
LanguageEnglish

Related product manuals