Pulse Oximetry (SpO2) Performance Product Specifications
29 - 38 V Series Operating Instructions
29.7 Pulse Oximetry (SpO
2
) Performance
V 12/V 21 in conjunction with:
A. VPS with Nellcor or Masimo
•SpO
2
Alarms
High Alarm Limits:
Low Alarm Limits:
B. VPS with Nellcor
1. Nellcor SpO
2
Sensor Compatibility
• OxiMax Adhesive Sensors: MAXAI, MAX-AL, MAXPI, MAXII, MAX-NI.
• OxiMax OxiCliq Sensors: OxiCliq A, OxiCliq P, OxiCliq I, OxiCliq N, OxiCliq Sensor Cable OC-3.
• OxiMax Reusable Sensors: DS-100A, D-YS, D-YSE, D-YSPD, Oxiband OXI-A/N, Oxiband OXI-P/I.
2. Nellcor SpO
2
Saturation Accuracy
NOTE: Measurement accuracy verification: The SpO
2
accuracy has been
verified in human experiments by comparing with arterial blood
sample reference measured with a CO-oximeter. Pulse oximeter
measurements are statistically distributed and about two-thirds of the
measurements are expected to come within the specified accuracy
range compared to CO-oximeter measurements.
• SpO
2
parameter minimum range is 70% to 100%.
• Invalid SpO
2
parameter displayed as (- -).
• SpO
2
parameter display resolution is 1%.
Adult: 75 to 101% in steps of 1%, default 101%
Pediatric: 75 to 101% in steps of 1%, default 101%
Neonate: 75 to 101% in steps of 1%, default 101%
Adult: 70 to 99% in steps of 1%, default 88%
Pediatric: 70 to 99% in steps of 1%, default 90%
Neonate: 70 to 99% in steps of 1%, default 92%
SENSOR ACCURACY
MAXAI, MAX-AL, MAXPI, MAXII, MAXNI (Adult), MAXNI (Neonate) 70% to 100% ± 2 digits
OxiCliq A, OxiCliq P, OxiCliq I, OxiCliq N(Adult), OxiCliq N (Neonate) 70% to 100% ± 2.5 digits
DS-100A, D-YS (Infant to Adult), OXI-A/N (Adult), OXI-P/I 70% to 100% ± 3 digits
D-YS with D-YSE Ear Clip, D-YS with D-YSPD Spot Clip 70% to 100% ± 3.5 digits
D-YS (Neonate), OXI-A/N (Neonate) 70% to 100% ± 4 digits