Display Options Dialog Display Setup
3 - 18 V Series Operating Instructions
3. Optional - If using a 12 lead set with the V 12, select the View All ECG button a second time to
view the next six (6) leads.
4. To return to the display mode, continue to select the View All ECG button until the display mode
displays.
3.2.20 Changing Lead/Size
The Lead/Size dialog provides a vehicle to change size of the ECG waveforms at the display when
using more than three ECG leads.
1. Select the Setup tab from the Navigation Area.
2. Select the Display button, then select the Lead/Size button.
The Lead/Size dialog displays.
3. Select the ECG lead by using the and scroll buttons. The leads available are based
on the lead set currently in use.
4. Select the lead size by using the and scroll buttons. The waveform sizes available
are: 0.125 cm/mV, 0.25 cm/mV, 0.5 cm/mV, 1 cm/mV, 2 cm/mV, 4cm/mV.
5. Select the Done button.
3.2.21 Touch Screen Calibration
NOTE: Recalibrate the touch screen whenever the touch interface becomes difficult to
maneuver.
1. Select the Setup tab from the Navigation Area.
2. Select the Display button, then select the Touch Screen Calibration button.
3. Touch and hold the areas where indicated. Release the area when indicated (as shown in FIGURE
3-7).

Table of Contents

Other manuals for Mindray V12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V12 and is the answer not in the manual?

Mindray V12 Specifications

General IconGeneral
BrandMindray
ModelV12
CategoryMedical Equipment
LanguageEnglish

Related product manuals