V Series Operating Instructions 29 - 37
Product Specifications Temperature Performance
Temperature Low Alarm Limits:
Delta Temperature High Alarm Limits:
The minimum measuring time required to obtain an accurate reading at the specified body site is
probe-dependent and should be indicated on the probe Instructions For Use.
For each temperature probe, information about the proper environmental conditions of use, storage,
and transport including temperature and humidity should be contained in the probe Instructions For
Use.
Adult: 25.0°C to 45.0°C in steps of 0.1°C, default Off
77.0°F to 113.0°F in steps of 0.1°F, default Off
Pediatric: 25.0°C to 45.0°C in steps of 0.1°C, default Off
77.0°F to 113.0°F in steps of 0.1°F, default Off
Neonate: 25.0°C to 45.0°C in steps of 0.1°C, default Off
77.0°F to 113.0°F in steps of 0.1°F, default Off
Adult: 15.0°C to 40.0°C in steps of 0.1°C, default Off
59.0°F to 104.0°F in steps of 0.1°F, default Off
Pediatric: 15.0°C to 40.0°C in steps of 0.1°C, default Off
59.0°F to 104.0°F in steps of 0.1°F, default Off
Neonate: 15.0°C to 40.0°C in steps of 0.1°C, default Off
59.0°F to 104.0°F in steps of 0.1°F, default Off
Adult: 0.0°C to 10.0°C in steps of 0.1°C, default Off
0.0°F to 10.0°F in steps of 0.1°F, default Off
Pediatric: 0.0°C to 10.0°C in steps of 0.1°C, default Off
0.0°F to 10.0°F in steps of 0.1°F, default Off
Neonate: 0.0°C to 10.0°C in steps of 0.1°C, default Off
0.0°F to 10.0°F in steps of 0.1°F, default Off