Temperature Performance Product Specifications
29 - 36 V Series Operating Instructions
29.6 Temperature Performance
The V 12/V 21 is capable of providing temperature measurements from two (2) channels for the 400
series probe.
V 12/V 21 in conjunction with:
• VPS or Temp module
1. Performance:
2. Temperature Alarms
The temperature function includes High and Low temperature alarms. Alarms can be triggered when
measured temperature is equal to or greater/lesser than the alarm setting.
Temperature High Alarm Limits:
Invalid Data: Invalid IBP Heart Rate data displayed as (- -).
Resolution: 1 BPM
Accuracy: ± 3 BPM or ± 3% at 30 to 250 BPM
± 5 BPM or ± 5% from 251 to 350 BPM
Range: Adult Mode:30 to 300 BPM
Scale: Selectable, Celsius or Fahrenheit
Range: 15°C to 45°C, inclusive
59°F to 113°F, inclusive
Operating Mode: Direct Mode
Tabular Trend: Yes
Graphic Trend: Yes
Invalid Data: Invalid Temperature data displayed as (- -).
Resolution: 0.1°C, 0.1°F
Scale for 400 Series Probes: Celsius or Fahrenheit
Range for 400 Series Probes: 15°C to 45°C (59°F to 113°F)
Accuracy: ± 0.1°C (15°C - 45°C), exclusive of probe errors
± 0.2°F (59°F - 113°F), exclusive of probe errors
Accuracy inclusive 400 Series Probes: ± 0.1°C (25°C - 45°C)
± 0.2°C (otherwise)
± 0.2°F (77°F - 113°F)
± 0.4°F (otherwise)
Probe Excitation: 400 series Probes: <200 µA, Thermister to Ground.
Delta Temperature range: 0°C to 10.0°C
0°F to 10.0°F
Delta T Resolution: 0.1°C
0.1°F
Transient Response Time 30 s